← All research

Mechanical ventilation vs bag-valve ventilation during cardiopulmonary resuscitation: a randomized crossover noninferiority trial.

Resuscitation plus · 2026-10-13

S Vinay, S Manu Ayyan, V T Amrithanand et al.

This single-center randomized crossover trial enrolled 104 intubated adults undergoing CPR who met termination-of-resuscitation criteria without return of spontaneous circulation. Among 99 completing both periods, mean per-participant median device-calculated inspired tidal volume was 368.0 mL with mechanical ventilation versus 514.4 mL with bag-valve ventilation. The adjusted difference was -146.2 mL (95% confidence interval, -187.0 to -105.5), failing the prespecified -150 mL noninferiority margin. Mechanical ventilation produced less variable volumes, but measurements used an unvalidated analyzer during a five-minute peri-termination window, and clinical outcomes were not assessed.

Clinical bottom line

Mechanical ventilation delivered lower device-calculated tidal volumes than bag-valve ventilation during CPR, but noninferiority was not demonstrated. These surrogate findings do not establish comparative clinical benefit.

Study limitation: An unvalidated analyzer, peri-termination population, and absence of clinical outcomes limit applicability to broader resuscitation care.

Read the original on PubMed

AI-generated from the published abstract with automated source checks. For research discovery; consult the original paper before applying findings.